SiCOMS® – Alarm Resetting

  • In case of a Pre-Alarm triggered:
    If the conditions for a Pre-alarm are not confirmed after pre-set time period, the Pre-Alarm will be reset automatically by the system.
  • In case of a Main-Alarm triggered:
    The Main Alarm will not be reset automatically by the system even after the conditions that led to it are not met any more. Main Alarm can only be reset manually at the SiCOMS®Evaluator by pressing the RESET button for 3 seconds.
  • In case of a Temperature Alarm triggered:
    Temperature Alarm is not being reset automatically by the system either! It must be reset by        pressing the RESET button at the Evaluator for 3 seconds
Warnung

Attention: Main-Alarm cannot be reset through the SiCOMS® / OCom Logger software, but only at the SiCOMS® Evaluator by pressing the RESET-Button for 3 seconds.