SiCOMS® – EEprom Parameters Failure

The display shows “CPU Error nnxx!”with   nn error code (refer to the following table) Error code Error description Help 01xx Failure within the CPU parameters Check CPU parameters (display brightness, …) 02xx Failure within the SiCOMS® parameters Check  SiCOMS® parameters (number of sensors, sensitivity, …) 04xx   Check manufacturing data If multiple errors are detected, the error code is calculated […]

SiCOMS® – EEprom Parameters Failure Read More »

SiCOMS® – Evaluator Failure

The display shows “OMD – Error nnnn!” with  nnnn:  error code (refer to the following table) Error code Error description Help 0001 No sensor detected • Check cables • Is the number of sensors set up correctly? 0002 Date and time invalid in real-time clock • Set date and time • If the error occurs again, replace Evaluator 0004 Failure

SiCOMS® – Evaluator Failure Read More »

SiCOMS® – Communication Failure

The display shows “Err.Sens.xx Stat: nn” with       xx: sensor number, for example 01            nn: error code (refer to the following table) Error code Error description Help 01 Sensor does not respond via interface “loop out” (connector 4) Restart system by pressing the ESC button 02 Sensor does not respond via interface “loop in” (connector 5) 03 Sensor does not

SiCOMS® – Communication Failure Read More »

SiCOMS® – Failure in SiCOMS® Sensors

The display shows “Err.Sens.xx: nnnn [text error message]” with       xx: sensor number, for example 01               nnnn: error code (refer to the following table) Error codes Error description Help 0001 must value for IR transmitter LED too high (> 3000) IR track must be cleaned and sensor must be recalibrated. If the error persists after recalibration, it means that

SiCOMS® – Failure in SiCOMS® Sensors Read More »

SiCOMS® – Troubleshooting

Error code and Error description If a failure is detected at the SiCOMS®, the green ready LED of the Evaluator turns off.  If there’s at least one SiCOMS® sensor online and functioning at this time, the green ready LED will blink as an indicator for reduced system operation. In this case, the system is still able to raise opacity alarm.

SiCOMS® – Troubleshooting Read More »

SiCOMS® – Data Sampling

As soon as a SiCOMS® Evaluator is powered up, it starts the initial self-check procedure and shows the “System Ready”  Status if all pre-set conditions are fulfilled. This self-check procedure is carried out in a certain cycle. Regardless of the green ready LED status of the SiCOMS® Evaluator, if working SiCOMS® sensors are connected, the SiCOMS® Evaluator samples the signals

SiCOMS® – Data Sampling Read More »